Iminodiacetic acid functionalised monolithic silica chelating ion exchanger for rapid determination of alkaline earth metal ions in high ionic strength samplesстатья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 апреля 2017 г.
Аннотация:Iminodiacetic acid has been covalently bonded to a bare silica monolith to produce the first reported high-performance monolithic chelating ion exchange column. Using the new column, separation and determination of traces of alkaline earth metal ions (low ppm) in high ionic strength samples (up to 2 M NaCl and KCl brines), could be achieved in under 40 s. At an eluent flow rate of 4 mL min-1 retention time precision was < 1.2% (n = 9) for Mg(II) and Ca(II), with detector linearity (n = 5) over the range 2-10 mg L-1 of between R2 = 0.985 and R2 = 0.995. In 1 M KCl and NaCl brine samples, detection limits of 0.2 mg L-1 were possible.